Defender 5000 Indicators
Zero Calibration is on the top of the list of
Calibration by default. Just press the Softkey
corresponding to the icon
Calibration.
Clear the pan and then press the Softkey
corresponding to the icon
The message 'Completed' will be displayed on
the screen.
Exit Zero Calibration by pressing the Softkey
corresponding to the icon
To return to the Main Menu, press the Softkey
corresponding to the icon

3.3.2 Span Calibration

Span Calibration uses one point. The span calibration point is established with a calibration mass
placed on the scale.
Note: Span Calibration should be performed after Zero Calibration.
Calibration procedures:
Long press the button
Menu.
Press the Softkey corresponding to the icon
enter the Calibration sub-menu.
Scroll to Span Calibration using the Softkey
corresponding to the icon
Press the Softkey corresponding to the icon
initiate Span Calibration.
Place a calibration mass of the specified weight on
the pan and press the Softkey corresponding to the
icon
. To change to a different calibration point,
input the value desired, and then place the
corresponding weight on the pan for calibration.
to initiate Zero
.
.
.
to enter the Main
to
.
to
EN-19