Advanced QC
The purpose of this procedure is to perform the advanced quality control checks built into the V.A.C.ULTA™ Therapy Unit
software.
These checks include:
•
•
•
•
•
•
•
•
Indicator Legend
Export Logs
Pressure Accuracy Testing
Pressure Calibration (if necessary)
Reset Counter
Pressure Transducer Check
V.A.C.® Therapy Alarm Testing
V.A.C. VERALINK™ Alarm Testing
Pump Performance Testing
The procedures for the V.A.C.ULTA™ Therapy Unit Advanced QC includes special
indicators that appear in this document but will not appear on the software screens
while using the device.
These indicators and their meaning appear below.
Graphic
1
A
green check
will appear when a test is successfully completed. If a
repeat the test. If the unit fails this second attempt, contact KCI (see page 16 of this
addendum).
Meaning
Do not use or press
Special attention required
Indicates which button on the screen matches the
applicable step in the instructions
red X
appears,
5