Unit Setup
Supplies and Equipment Needed
calibrated digital manometer
vacuum source
locally sourced flexible tubing that fits over the canister tubing and the manometer barb
two KCI 500mL canisters (KCI PN M8275063 or M8275071). Orderable PNs: M8275063/5 and M8275071/5
These are standard canisters that can be reused for up to six months as a test canister.
On one canister, block the sensing port with tape; this will simulate a full canister.
Leave the sensing port open on the other canister. Write on both canisters 'Test Canister-Not For
Human Use' and the in-use date.
one KCI V.A.C. VERALINK™ test cassette (KCI PN SE-000387). Orderable PN: SE-000387/05
This cassette can be reused for up to four months as a test cassette. Write on cassette 'Test Cassette-
Not For Human Use' and the in-use date.
one KCI V.A.C. VERAFLO™ bracket test fixture (KCI PN SE-001179). Orderable PN: SE-001179/05
This fixture can be reused for up to 4 months as a test fixture. Write on fixture 'Test Fixture - Not For
Human Use' and the in use date
1. Disconnect the V.A.C.ULTA™ Therapy Unit from the power supply.
2. Remove the bellows.
3. Connect the unit to the test fixture assembly and manometer.
3