4. Press the
Arrow
5. Verify the manometer reading is +/- 10mmHg of 125 mmHg and record the
number on the V.A.C.ULTA™ Therapy Unit Checklist (included in the V.A.C.ULTA™
Therapy System Maintenance manual).
6. The unit has passed the test and a green check mark will be displayed in the
Status box. If the unit fails the Pressure Calibration screen will appear
automatically.
7. Disconnect the manometer
8. Reinstall the bellows.
9. Continue to the V.A.C.® Therapy Alarm Testing section
button displayed next to
Pressure
Accuracy.
10