Tools and Supplies
•
•
•
Two KCI 500mL Canisters
Preventative Maintenance
The purpose of this procedure is to check instill pump performance and pressure accuracy for the V.A.C.ULTA™ Therapy
Unit software.
To ensure accuracy, this procedure is required every six months.
two KCI 500mL canisters (KCI PN M8275063 or M8275071). Orderable PNs: M8275063/5 and M8275071/5
•
These are standard canisters that can be reused for up to six months as a test canister.
On one canister, block the sensing port with tape; this will simulate a full canister.
•
Leave the sensing port open on the other canister. Write on both canisters 'Test Canister-Not For
Human Use' and the in-use date.
one KCI V.A.C. VERALINK™ test cassette (KCI PN SE-000387). Orderable PN: SE-000387/05
•
This cassette can be reused for up to four months as a test cassette. Write on cassette 'Test Cassette-
Not For Human Use' and the in-use date.
one KCI V.A.C. VERAFLO™ bracket test fixture (KCI PN SE-001179). Orderable PN: SE-001179/05
•
This fixture can be reused for up to 4 months as a test fixture. Write on fixture 'Test Fixture - Not For
Human Use' and the in use date
Sensing Port with Tape
Test Cassette
Test Fixture
2