KCI V.A.C.ULTA Update-Anleitung Seite 12

Vorschau ausblenden Andere Handbücher für V.A.C.ULTA:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 17
If the Target pressure box numbers and pressure reading on the
manometer do not match use the + or – button to adjust the display
to match.
6. Press the
and an offset value will be generated.
7. Press the
8. Press the
9. Disconnect the Pressure Calibration Setup.
10. Reconnect unit for testing based on the Preventive Maintenance/Unit Setup
instructions.
11. Press the
QC Checklist
12. Press the
to the
V.A.C. ® Therapy
13. Repeat the Pressure Accuracy Check.
If the calibration can not be completed contact KCI for next steps
(see the Customer Contact Information section page 16 of this
addendum).
14. Disconnect the manometer.
15. Resinstall the bellows.
Play
button next to the Measured box for the Target Pressure
Set
button to save the 125 calibration point.
QC Checklist
button to continue.
Advanced QC
button on the
screen.
Arrow
button displayed next to
screen.
V.A.C. ® Therapy
to advance
12