1. Description
General
Tec 3 Vaporizers are designed for 'out of circuit' use in
continuous flow techniques of inhalation anaesthesia.
The vaporizers are temperature and flow compensated so
that their output remains relatively constant despite
cooling due to vaporization and variations in inlet flow (see
- 'Effects of Variables').
Each vaporizer is clearly labelled with the name of the
anaesthetic agent for which it is designed.
Control Dial
A single control dial with a concentration scale calibrated
in % of anaesthetic agent vapour per total volume (v/v) is
employed to set the desired concentration of the
anaesthetic agent.
To prevent accidental displacement of the control dial from
the 'OFF' position to 'ON', a release button is incorporated
in the dial assembly. Simultaneous depression of the
release button and anti-clockwise rotation of the dial is
necessary to achieve an 'ON' setting.