Warnings
DO NOT FILL THE VAPORIZER WITH ANY
ANAESTHETIC AGENT OTHER THAN THE ONE
SPECIFIED ON THE FRONT LABEL. THE VAPORIZER IS
DESIGNED FOR THAT AGENT ONLY. ANY OTHER
AGENT THAN THAT SPECIFIED CAN PROVE TO BE
DANGEROUS TO A PA'HENT.
DO NOT CARRY TOE VAPORIZER BY THE CONTROL
DIAL.
HANDLE WITH CARE.
DO NOT MODIFY, TAMPER WITH OR DISASSEMBLE
THE VAPORIZER BECAtJSE OF THE DANGERS OF
DAMAGING THE UNIT AND ALTERING THE ACCURACY
OF GRADUA'nON.
DO NOT IMMERSE THE VAPORIZER IN ANY UQUID,
INCLUDING WATER.
DO NOT STERILISE THE VAPORIZER.
DO NOT DRAIN THE ANAESTHETIC AGENT INTO ANY
CONTAINER OTHER THAN A PROPERLY MARKED
CONTAINER.
DO NOT FILL "niE VAPORIZER UNLESS raE CONTROL
DIAL IS IN THE 'OFF' POSITION.
DO NOT TURN THE DIAL 'ON' DURING FILLING OR
ATTEMPT TO FILL BEYOND THE 'FULL' MARK.
TO AVOID POTENTIAL LEAKS, BEFORE MOUNTING THE
VAPORIZER ON THE SELECTATEC MANIFOLD, CHECK
THAT THE MANIFOLD PORT VALVE 'O' RINGS ARE
INTACT AND THAT THERE IS NO FOREIGN MATTER
AROUND THE MATING SURFACES.
KEEP THE VAPORIZER UPRIGHT AT ALL TIMES.
DO NOT PUT WATER OR ANY OTHER SOLVENT IN A
VAPORIZER
Cautions
Do not tiave the vaporizer serviced by anyone other ttian
an Ohmeda Authorised Service Centre.
Turn the Vaporizer 'off' when not in use.
Before filling vaporizers with screw cap fillers, if the
vaporizer is pressurised turn the screw cap slowly.
Before filling keyed filler vaporizers, if the unit is
pressurised turn the screw slowly when removing the
plug.