6. Operating Instructions
WARNING
KEEP THE VAPORIZER UPRIGHT AT ALL TIMES
DO NOT CARRY THE VAPORIZER BY THE CONTROL
DIAL
HANDLE WITH CARE
T u m i n g On
To turn the vaporizer 'ON' depress the control dial release
button.
To avoid the inadvertent delivery of small concentrations
the control dial should be turned to 'OFF' when the
vaporizer is not in use.
Filling a n d Draining
General
WARNING
DO NOT FILL VAPORIZER WFTH ANY AGENT OTHER
THAN THE ONE SPECIFIED ON THE FRONT LABEL. I H E
VAPORIZER IS DESIGNED FOR THAT AGENT ONLY.
ANY OTHER AGENT THAN THAT SPECIFIED CAN
PROVE TO BE DANGEROUS TO A PA'HENT.
DO NOT FILL VAPORIZER UNLESS THE CONTROL DIAL
IS IN THE 'OFF' POSITION.
DO NOT TURN THE DIAL 'ON' DURING FILLING OR
ATTEMPT TO FILL BEYOND THE 'FULL' MARK.
DO NOT DRAIN THE AGENT INTO ANY CONTAINER
OTHER THAN A PROPERLY MARKED CONTAINER.
Periodically check the agent level. The vaporizer should be
refilled at appropriate intervals. The vaporizer v\rill function
satisfactorily as long as there is agent visible in the sight
glass.
The vaporizer should be filled and used in an upright
position. Small deviations from the upright position will not
affect the output or the safety of the vaporizer, but because
the agent depth is shallow in relation to the diameter of the
vaporizing chamber more frequent checks of the agent
levels should be carried out when small deviations from the
upright position occur in order to avoid obtaining a
misleading impression of the amount of agent in the
vaporizer.
It is good practice when the level is low, to drain the
vaporizer of residual anaesthetic agent. The small amount
of agent should be discarded.
At intervals - ideally not exceeding two weeks - the
vaporizer should be drained into an appropriately marked
container when the agent level is low and the agent
discarded. Less frequent intervals may be used when the
anaesthetic agent does not contain additives or stabilising
agents.
Tuming On
Einschalten
Mise en Marche
Turning On