14 | Serielle Kommunikation
Setup
Read/write
instrument serial
Read
instrument S/W
version
Read/write
device serial
Read/write
device type
Read device
S/W version
Read device
S/W bootloader
version
Measure [Value read
request]
Set Main Run Time
Read/write real time
clock
52 / 82
Geräte-Setup lesen und schreiben
Lesen
Daten-
Antwort
sequenz
rDis\n
[S/N]
rDiv\n
[SW ver. No.]
rDds\n
[S/N]
rDdt\n
[Device type],
e.g P60
rDdv\n
[SW ver. No.]
rDdb\n
[SW B ver.
No.]
rDmv\n
[Same digit as
on the screen]
-
-
rDcl\n
[yyyy-MM-dd
HH:mm:ss]
Write
Daten-
Wert/Text
sequenz
wDis [text]\n
S/N
-
-
wDds [text]\n
S/N
wDdt [value]\n 0 - 9
-
-
-
-
-
-
wD\r [value]\n
In minutes.
wDcl [text]\n
yyyy-MM-dd
HH:mm:ss
Sensistor_Sentrac_Strix_Betriebsanleitung_ninb65de1-01-(2011)
INFICON
Erklärung
Max 8 Zeichen
X.XX.XX
Gerät z.B. P60.
Max 8 Zeichen
0 = UNBEKANNT
1 = P60
2 = PL60
3 = COMBOX60
4 = TGF11
5 = AP29_SENTRAC
6 = P50
7 = AP29
8 = COMBOX
9 = STRIX
X.XX.XX
X.XX.XX
Im Terminal schreiben
Sie "wD\\r 60\n" für 1
Stunde.