• Insert the key-bolts (9) from above into the hinges
• Align the door by fastening the hinge screws (7a) in the elongated holes of the hinge. The
door must be set straight and the locking must operate smoothly.
3.
Spare parts list
Title
Locking block (for incubators delivered before October 1999)
Rubber profile (narrow) for 4-fold divided inner glass door CB 150
Rubber profile (narrow) for 6-fold divided inner glass door CB 210
Individual glass door of 4-fold divided inner glass door CB 150, gas proof, complete 8010-0051
Individual glass door of 6-fold divided inner glass door CB 210, gas proof, complete 8010-0035
Rubber profile for individual glass door of 4-fold divided inner glass door CB 150
Rubber profile for individual glass door of 6-fold divided inner glass door CB 210
4.
Service
In case of condensation on the inner face of the divided inner glass door, the Offset of the door
heating should be increased by 1°C. If this does not prove sufficient, it is possible to increase
the Offset further by steps of 0.5°C. Consult the BINDER service to adjust the setting.
BINDER telephone hotline:
BINDER fax hotline:
BINDER e-mail hotline:
BINDER service hotline USA: +1 866 816 8191 (toll-free in the USA)
BINDER Asia Pacific:
BINDER Internet homepage
BINDER address
International customers, please contact your local BINDER distributor.
Installation of divided inner glass door 01/2009
Key-bolt (9) for hinge
Hinge (7)
with set-in divided inner glass door
+49 (0) 7462 2005 555
+49 (0) 7462 2005 93555
service@binder-world.com
+ 603 6204 2855
http://www.binder-world.com
BINDER GmbH, post office box 102, D-78502 Tuttlingen
Part-No.
6002-0124
6005-0102
6005-0103
6005-0100
6005-0101
Seite/page/ página 10/22