PJX Series Balance
3. To change the calibration mode, press No until SpaN
(span calibration) is displayed.
4. The Calibration mass value will be shown in the screen.
After the display shows 200.0000 g, place weight(s) of
200 g on the pan for calibration. To change to the
calibration point of half full capacity (e.g. 100 g), press
the Function / Mode button. After the screen shows
"Place weights" and "100.000 g", place weight(s) of 100
g on the pan for calibration.
5. Once the span calibration is completed successfully,
CAL.Done will display.
Press any button to return to the previous screen.
6. Remove the weight(s), and the reading will be set to
zero.

5.2.5 Linearity Calibration

Linearity calibration uses three calibration points, one at zero load and the others at specified loads.
With no load on the balance, press Linearity Calibration to begin the process.
The balance captures the zero point, and then prompts for the next weight.
Continue to follow the instructions on the display until the calibration is completed.
Steps for linearity calibration
1. Long press the Tare / Menu-Cal button, and the
Calibration Menu will display.
EN-21