Linearity Calibration - OHAUS Pioneer PX Series Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

EN-26
2. Press Yes to enter the Calibration Menu. To change
the calibration mode, press No until Span Cal (span
calibration) is displayed.
3. Press Yes to begin the span calibration.
4. The calibration masses value will be shown in the
screen. After the display shows "Place weights" and
"100.000 g", place weight(s) of 100 g on the pan for
calibration. To change to the calibration point of half full
capacity (e.g. 50 g), press the Function / Mode button.
After the screen shows "Place weights" and "50.000 g",
place weight(s) of 50 g on the pan for calibration.
5. Remove the weight from the pan.
6. Once the span calibration is completed successfully,
"Calibration done" will display.
Press any button to return to the previous screen.

5.2.5 Linearity Calibration

Linearity calibration uses three calibration points, one at zero load and the others at specified loads.
For detailed calibration mass information, please see the specification tables in section 9.1.
With no load on the scale, Linearity Calibration can be performed.
The balance captures the zero point, and then prompts for the next weight.
PX Series Balance