PR Series Balance
Steps for span calibration
1. Press and hold the Tare / Menu-Cal button, and the
Calibration Menu will display.
2. Press Yes to enter the Calibration Menu.
3. To change the calibration mode, press No until SpaN (span
calibration) is displayed.
4. The calibration mass value will be shown in the screen. After
the display shows 200.0000 g, please place weight(s) of 200
g on the pan for calibration. To change to the calibration
point of half full capacity (e.g. 100 g), press the Function /
Mode button. After the screen shows "Place weights" and
"100.000 g", place weight(s) of 100 g on the pan for
calibration.
5. Once the span calibration is completed successfully,
CAL.Done will display.
Press any button to return to the previous screen.
6. Remove the weight, and the reading will be set to zero.

5.2.5 Linearity Calibration

Linearity calibration uses three calibration points, one at zero load and the others at specified loads.
With no load on the balance, press Linearity Calibration to begin the process.
EN-15