EN-8

Auto Shut-off

When Auto Shut-off is activated, the balance will shut off after three minutes of
non use.
1. When the balance displays
or Off.
2. To change the displayed setting, press Off.
3. To accept the displayed setting, press On. The display will again indicate
.
4. Change to another menu item, or exit as described above.

Activating Units

1. Start with the balance OFF. Press On until
2. Press Mode until
3. With
current setting of
4. To change the displayed setting, press Mode.
5. To accept the displayed setting, press On. The display will advance to the
next weighing unit.
6. Repeat steps 4 and 5 for each weighing unit and parts counting. When the
last unit has been accepted, the display will again indicate
7. Repeatedly press Mode to change to another menu item or to exit, press
Mode until

Linearity Calibration

Calibration masses are available as accessories for performing this procedure.
2 Text new.pmd
, press On to view the current setting On
is displayed.
displayed, press On to display the grams indicator "g" with its
or
.
g
g
is displayed, then press On to store settings.
Balance
Required Calibration Masses
JS10
5g, 10g
JS500
200g, 400g
8
Sapphire JS10 and JS500
is displayed.
.
6/10/2004, 9:53 AM