Technical Data
DC Current
Measurement Range
Accuracy
500,00 mA
± (0,15% rdg. + 20 D) 0,15 mV/mA
5000,0 mA
± (0,1% rdg. + 20 D)
50,000 mA
± (0,15% rdg. + 10 D) 3,3 mV/mA
500,00 mA
± (0,1% rdg. + 20 D)
5,0000 A
± (0,5% rdg. + 10 D)
10,000 A*
± (0,5% rdg. + 20 D)
* 10 A continuous, 20 A for max. 30 s including 5 min cooling down
AC, AC/DC Current
Range
Hexagon 710
50...60 Hz
500,00 mA
5000,0 mA
50,000 mA
± (1,0% rdg. + 40 D) ± (0,5% rdg.
500,00 mA
5,0000 A
10,000 A*
40Hz..1kHz
500,00 mA
5000,0 mA
50,000 mA
± (1,0% rdg. + 40 D) ± (0,7% rdg.
500,00 mA
5,0000 A
10,000 A
1kHz..10kHz
500,00 mA
5000,0 mA
50,000 mA
not specified
500,00 mA
5,0000 A
10,000 A
* 10 A continuous, 20 A for max. 30 s including 5 min cooling down
74
manuals search engine
Apparent Voltage
Overload Protection
0,15 mV/mA
0,63 A / 500 V (710)
1 A / 600 V (720)
3,3 mV/mA
0,03 V/A
12,5 A / 500 V (710)
0,03 V/A
15 A / 1 kV (720)
Hexagon 720
Apparent Voltage Overload Protection
0,15 mV/mA
0,63 A / 500 V
0,15 mV/mA
(Hexagon 710)
3,3 mV/mA
1 A / 600 V
+ 50 D)
3,3 mV/mA
(Hexagon 720)
0,03 V/A
12,5 A / 500 V (710)
0,03 V/A
15 A / 1 kV (720)
0,15 mV/mA
0,63 A / 500 V
0,15 mV/mA
(Hexagon 710)
3,3 mV/mA
1 A / 600 V
+ 50 D)
3,3 mV/mA
(Hexagon 720)
0,03 V/A
12,5 A / 500 V (710)
0,03 V/A
15 A / 1 kV (720)
0,15 mV/mA
0,63 A / 500 V
± (2,0% rdg.
0,15 mV/mA
(Hexagon 710)
+ 50 D)
3,3 mV/mA
1 A / 600 V
3,3 mV/mA
(Hexagon 720)
not specified
0,03 V/A
12,5 A / 500 V (710)
not specified
0,03 V/A
15 A / 1 kV (720)