Bedienungsanleitung_MC_ALU_DE_EN_RevD.docx
BRIEFOPERATIONGUIDE
M
C
ULTI
ONTROL
Loading
1. Insert the CertoClav bottom plate into the sterilization chamber.
2. Fill the chamber with 1-1.5 liters of distilled water.
For CertoClav Essential, add 1/3 tap water
3. Place the items to be sterilized on the bottom plate.
Closing and locking
1.
Check if the gasket is inserted in the lid.
2.
Turn the orange knob to "OPEN".
3.
Place the lid on the chamber and turn it clockwise until the handles are in line with each other.
4.
Turn the orange knob to position "CLOSE" for locking the lid.
5.
Connect a tube to the steam release cock, usual it is opened and sealed.
6.
Plug in the power cable. Plug in the heater cable into the pot and controller. Plug in the temperature sensor
cable and screw it into the controller.
Starting a program
Select a sterilization program in the control unit and push start.
Venting
If venting is activated, then the sterilization chamber is vented at 115°C for 10 minutes.
Sterilization
In case of sterilization times longer than 60 minutes, close the black steam release cock after the
sterilization temperature has been reached. Turn knob clockwise.
Opening
Wait until the temperature of the Chamber is below 80°C. Turn red knob to OPEN and open the lid of
autoclave. If liquids have been sterilized, wait until the chamber temperature drops below 70°C°. Otherwise
there is the risk of burns from delayed boiling.
Sterilization timetable
If you want to choose a new program in the control unit, consider the following sterilization holding time
for different volume of liquid media.
Volume of loading
liquid
3000 ml
2000 ml
1000 ml
500 ml
CertoClav Sterilizer GmbH
Peintner Strasse 10
A-4060 Leonding / Austria
www.certoclav.com
Alu
Sterilization
time until 121°C
30 Minutes
25 Minutes
20 Minutes
15 Minutes
support@certoclav.com
Page 43 of 44