Herunterladen Inhalt Inhalt Diese Seite drucken

Cleaning And Maintenance; Error Messages - CertoClav MultiControl 12 Bedienungsanleitung

Laborautoklav
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Bedienungsanleitung_MC_ALU_DE_EN_RevD.docx

Cleaning and maintenance

If necessary, the controller can be wiped with a moist, not wet, cloth. Maintenance must only be done by
the manufacturer or an authorized body. No regular maintenance necessary.

Error messages

Error messages are shown
on the display as follows:
Message
"Program stopped"
"Power loss"
"Delta-T error"
"Sensor error"
"Heater error"
"Profiler error"
Display background light
is turned off permanently
Other errors
CertoClav Sterilizer GmbH
Peintner Strasse 10
A-4060 Leonding / Austria
www.certoclav.com
Possible reason
The program has been stopped
manually by user.
Program has been stopped because of
power loss.
The allowed temperature range of T-
1/+2°C has been exceeded during
holding time.
Error in temperature sensor.
Maximum allowed temperature
exceeded (143°C) OR target
temperature is not reached.
Fuses defective
support@certoclav.com
Page 40 of 44
Troubleshooting
Let the autoclave cool down before you start
again.
Ensure that the fuse holder has been
assembled tightly and that the power cable is
mounted correctly.
Check if the temperature cable has been
plugged and screwed correctly into the
controller. Check if the autoclave chamber is
not out of water.
Most often this means the sensor is not
plugged in into controller. Other possible
reason: cable break.
Please check if temperature sensor is working
properly or autoclave chamber is not out of
water.
Please navigate to Settings -> Factory reset
and press the OK button. This is necessary to
solve software issues.
Please check if the power switch is turned on.
Please replace both 5mm 10A slow acting
fuses placed at the shelf of the black power
connector.
Navigate to Settings -> Factory reset and
press the OK button. This is necessary to
solve software issues.