Bedienungsanleitung_MC_ALU_DE_EN_RevD.docx
Description
Display contrast is bad
Autoclave runs out of
water during sterilization
Water is dropping out of
the gasket of the lid
If the errors cannot be resolved contact your dealer or the manufacturer
E-Mail:
Usage of protocol printer
After connecting a CertoClav protocol printer (Art. No. 1300187) the following protocol is printed:
Header
10
Type of autoclave
15
Program name
30
Cycle number
Program steps
SxxA
Heat up
SxxH
Hold
SxxC
Drying
SxxE
End
Footer
40
Control message.
Sterilization successful?
CertoClav Sterilizer GmbH
Peintner Strasse 10
A-4060 Leonding / Austria
www.certoclav.com
Possible reason
Lid gasket defective
Lid gasket insert wrong
Autoclave not locked correct
Steam release cock is open
...
On heat-up, the gasket of the lid will
be compressed. If the gasket is wet
also, this can cause a few drops
coming out of the gasket during heat-
up.
support@certoclav.com
support@certoclav.com
Troubleshooting
Navigate to Settings -> Contrast and press the
OK button until the contrast is well.
Check if the gasket has been placed correctly
into the lid. The text "Place This Side Facing
Down" on the gasket must be readable for
the user. Check if the red knob on the handle
is in locking position during sterilization. For
cycles with sterilization time longer than 60
minutes, close the steam release cock.
If the dropping doesn' t stop after heat-up,
please remove the gasket and clean it with
distilled water. If the problem is not resolved,
replace the lid gasket.
/ Phone: +43 (0) 732 / 674 278
Page 41 of 44