Herunterladen Inhalt Inhalt Diese Seite drucken

I-Tech MIO-SONIC Benutzerhandbuch Seite 43

Ultraschall-therapie
Vorschau ausblenden Andere Handbücher für MIO-SONIC:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

device will enter in standby mode. If the warming feature is not
needed, press both the MODE button and the TIME button
simultaneously, in this way the device will go back to standby mode.
When the device is in standby mode, the modulation duty cycle is
defaulted at 5% and the (L) indicator light will be illuminated.
To restart the warming feature, you will have to power off the device
and turn it back on again.
3.
Wash the area to be treated so that it is free
of oil and dirt. Apply a generous layer of
ultrasound transmission gel on the treatment
area. The gel is fundamental to ensure a
correct fit between the treated area and the
ultrasound head and so the efficiency of the therapy.
Press several times the MODE button to
4.
select the modulation duty cycle. The mode
button has three levels, Low (L, preset),
Medium (M) and High (H).
5.
Press several times the TIME button: the lights
will light up in seuqnce related to the 5-10-15
minutes of therapy and the device will start
working. Set the time, the device will start
automatically the treatment.
6.
Place the ultrasound head on treatment area: is
important
continuously and uniformly the ultrasound
head around the treatment area, with slow
and circular or vertical movements of at least
7-8
cm.
IACER Srl
DO NOT APPLY THE ULTRASOUND HEAD TO THE SKIN OF
THE PATIENT during the warming period! The device will
automatically exit the head warming feature if any load is
detected in the preheating process.
perform
PROHIBITED
therapy
moving
MAINTAIN
THE
39
MNPG71-04

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis