Locksystem
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

component have to be met.
A suitable tool (i.e. a torque
wrench) has to be used to set
the locking torque accordingly.
• Before the prothesis is han-
ded over to the user, all mo-
dular components have to be
secured against accidental
loosening or twisting with
suitable thread-locking fluid.
• Safety related regulations for
individual components (e.g.
specific combination possibi-
lities, maintenance intervals
etc.) must be followed.
• If different components (from
one or from several manu-
facturers) are combined: Only
use components which meet
the requirements
of DIN EN ISO 10328 and the
Medical Devices Act.
• In the event of damage to a
prosthesis which combines
components from different
manufacturers, every ma-
nufacturer can only be held
liable for the failure of those
components produced by
WJT 8 Lock System
them. Beyond that, manu-
facturers can only be held
liable if their components can
be proven to have caused
damage to or failure of com-
ponents which were produced
by other manufacturers.
• Always observe maximum load!
• Overloading the components
may cause the prosthesis to
malfunction and will repre-
sent a safety risk for the user.
• Do not shorten the pins of the
lock systems!