• The prosthesis and its functional
components must be checked for
proper function, possible wear
and tear and potential damage
by an orthopedic technician at
regular intervals.
• If the prosthesis or any of its
components have been subject
to disproportionate stress (e.g.
due to a fall), they must be che-
cked for potential damage by
an orthopedic technician before
further use.
• This product is intended for use
on one person. It may not be
reused on another person.
When using WJT lock systems
components to fabricate a
prosthetic limb, the following
aspects have to be considered:
• The reinforcement should
match the load the prosthesis
has to carry and the user's
activity level. In any case the
orthopedic technician has
to perform an individual risk
assessment for the user in
question. Where necessary,
EN-4
the stability of the lamination
User Manual
will have to be increased with
additional lamination materials.
• Use lock systems only accor-
ding to their intended purpose.
• If lock systems are combined
with modular components
with different weight limits,
the weight limit of the we-
akest component applies to
the whole prosthesis.
• Using individual, certified
components with the
CE-marking does not release
technicians from their duty
to check component combi-
nations within their means
for fitness of purpose, proper
assembly and safety.
• Should there be evidence
that a certain combination
of components is not as safe
as required, the components
must not be combined.
• The prosthesis has to be
configured according to the
recognized technical rules of
the trade.
• The locking torque specifica-
tions given for the modular