9. Faults

9. Faults
Personnel: Skilled personnel
Protective equipment: Protective gloves
For contact details see chapter 1 "General
information" or the back page of the operating
EN
instructions.
WARNING!
Physical injuries and damage to property
and the environment caused by high
temperatures
In the event of a defect, extreme temperatures
can be present at the instrument.
Wear the requisite protective equipment.
Error
No function
The calibrator does not respond
when the power cord is
connected and the switch is on.
The fuse tripped as the
instrument was switched on
Final temperature is not reached
The display is working properly
but the temperature does not
increase and the calibrator
displays error code Ht.
The display is working properly
but the temperature does not
increase and the calibrator
indicate Err after few seconds.
The display shows a different
temperature than is present in the
block
The temperature exceeds the set
point
The calibrator is barely cooling
down
810 or 786 appears on the display. Internal probe is faulty
24
Causes
The voltage supply is not established correctly
The fuse is defective
The power cord is defective
The main switch is faulty
The fuse is wrong
The main switch is faulty
Short-circuit in the heating element
The solid-state relay or the heating element is
faulty
The overtemperature switch has tripped
The safety thermostat has been triggered for
over temperature
The heater is faulty
The temperature controller is faulty
The internal probe is not calibrated
The controller is defective
The control circuit board is defective
The controller is defective
Fan is defective
CAUTION!
Physical injuries and damage to property
and the environment
If faults cannot be eliminated by means of the
listed measures, the instrument must be taken
out of operation immediately.
Contact the manufacturer.
If a return is needed, please follow the
instructions given in chapter 11.2 "Return".
Measures
Check voltage supply
Replacing the fuse
Replace the power cord with an equivalent
Send in for repair
Check that the proper fuse is inserted for
the supply voltage and replace the fuse, if
necessary.
Send in for repair
Send in for repair
Verify the set point of the safety thermostat Set:
Ref to the level 4 of the menu.
Press any button to see if restore the heating.
Send in for repair
Replace the temperature controller
Return the temperature dry-well calibrator for
calibration
Send in for repair
Send in for repair
Send in for repair
Send in for repair
WIKA operating instructions, model CTD4000