Autoclave for more than 20 min. at 121°C, or 15 min. at 132°C, or 3 min. at 134°C.
3) The handpiece should remain in the autoclave pouch until required for use.
CAUTION
Do not autoclave the product with other instruments even when it is in a pouch. This is to prevent
possible discoloration and damage to the product from chemical residue on other instruments.
Keep the product in suitable atmospheric pressure, temperature, humidity, ventilation, and sunlight.
The air should be free from dust, salt and sulphur.
Immediately after use, the product should be cleaned, lubricated and sterilized. If blood remains on
the external or internal surfaces it can become clotted and cause rust.
Do not heat or cool the product too quickly. Rapid change in temperature could cause damage to the
product.
If the sterilizer chamber temperature may exceed 135°C during the dry cycle then delete the dry
cycle.
Autoclave sterilization is recommended for the product. The validity of other sterilization methods is
not confirmed.
Do not touch the product immediately after autoclaving as it will be very hot and must remain in a
sterile condition.
NOTICE
NSK recommends Class B sterilizers as stated in EN13060.
5. Periodical Maintenance Checks
Perform periodical maintenance checks every three months, referring to the check sheet below. If any
abnormalities are found, contact your Authorized NSK Dealer.
6. Symbol
This product is Autoclavable up to Max.135°C.
This product can be washed via Thermo Disinfector.
Conforms to CE European Directive of "Medical device directive 93/42/EEC."
Manufacturer.
Authorized representative in the European community.
Caution: U.S. Federal law restricts this device to sale by or on the order of a licensed physician.
7. Warranty
NSK products are warranted against manufacturing errors and defects in materials. NSK reserves the right
to analyze and determine the cause of any problem. Warranty is voided should the product be not used
correctly or for the intended purpose or has been tampered with by unqualified personnel or has had non
NSK parts installed. Replacement parts are available for seven years beyond discontinuation of the model.
8. Option Parts List
9. Disposing product
In order to avoid the health risks of operators handling the disposal of medical equipment, as well as the
risks of environmental contamination caused thereof, a surgeon or a dentist is required to confirm the
equipment is sterile. Ask specialist firms who are licensed to dispose of specially controlled industrial
wastes, to dispose the product for you.