Air Motor (FX205 M4/FX205m M4/FX205 B2/FX205m B2)
CAUTION
Read this Operation Manual before use to fully understand the product functions and file for future
reference.
When operating the product always consider the safety of the patient.
This product does not consider patient's age, gender nor physical condition.
This product does not consider operator's age, gender nor physical condition.
Users are responsible for the operational control, maintenance and continual inspection of this
product.
Do not attempt to disassemble the product nor tamper with the mechanism except as recommend
by NSK in this Operation Manual.
Do not allow any impact on to the product. Do not drop the product.
Operators and all others in the area must wear eye protection and a mask when operating this motor.
Should the product function abnormally, cease operation immediately and contact your Authorized
NSK Dealer.
Do not use high acid water or sterilizing solutions to wipe, immerse or clean the product.
The products are delivered in a non-sterile condition and must be autoclaved prior to use.
Perform regular function and maintenance checks.
If the product is not used for a long period check it is functioning correctly before using on a patient.
To avoid clinical downtime it is recommended that a spare be kept on hand in case of a breakdown
during surgery.
U.S. Federal law restricts this device to sale by or on the order of a licensed physician.
2
1. Setting of Air & Water Supply Pressure
Measure the supply pressure at the handpiece / hose connection point and set the pressure to the value
specified on the specification table. (Fig. 1)
For Multi Gauge information refer to Option Parts List.
WARNING
Do not exceed the optimum pressure specified on the specification table.
CAUTION
Do not use air contaminated by dust, moisture and oil.
2. Connection & Disconnection of the Handpiece
2-1 Connection
1) Connect the handpiece direct to the motor. (Fig. 2)
2) Make sure the handpiece is firmly connected to the motor.
2-2 Disconnection
Hold the motor and the handpiece separately and disconnect with care.
CAUTION
Disconnect the handpiece only after the motor has completely stopped rotating.

3. Check before treatment

Connect the handpiece to the motor and check for motor vibration, noise and overheating. If any
abnormalities are found do not use the motor and contact your Authorized NSK Dealer.