Remove the bur only after the handpiece has completely stopped rotating.
Do not rotate the handpiece when the Bur Lock Ring is turned to "
not inserted.
Always keep the bur shank clean.Entry of debris into the chuck, via the bur shank, could cause bur
rotation slip and also prevent the bur from being securely located in the chuck.
Do not exceed the bur speed recommended by the bur manufacturer.
Do not exceed maximum bur length recommended by the handpiece manufacturer.
Do not apply excess pressure to the bur as it may break or bend or become difficult to remove.
DO NOT use burs with problems listed below as the bur may break or disengage from the chuck.
- Bent, deformed, anisomerous (worn), rusted, broken, deficient bur.
- Bur which is cracked on the edge or axis.
- Non-ISO standard, or tampered bur.
3. Check before treatment
Check for handpiece vibration, noise and overheating. If any abnormalities are found do not use the
handpiece and contact your Authorized NSK Dealer.
4. Maintenance
After each patient maintain the product as follows.
4-1 Cleaning
1) Remove dirt and debris from the product. Do not use a wire brush.
2) Wipe clean with alcohol-immersed cotton swab or cloth.
This icon denotes that the product can be washed via Thermo Disinfector. (FX65m)
Refer to the Thermo-Disinfector manual.
12
CAUTION
" (OPEN) position or a bur is
After washing with Thermo-Disinfector and prior to lubrication, dry the product until all internal
moisture is thoroughly removed. Thermo-Disinfector moisture remaining inside the product could
reduce the effect of lubrication and could cause corrosion inside of the product.
To clean the product never use any solvent such as benzine or thinner.
4-2 Lubrication
NSK PANA SPRAY Plus
Apply NSK PANA SPRAY Plus every time after each use and/or before autoclaving.
1) Remove the bur from the handpiece.
2) Insert the Spray Nozzle into the Spray Port nozzle on the can.
3) Insert the Spray Nozzle in rear of the handpiece. Hold the handpiece and spray for approximately 2-3
seconds. Apply lubricant until it expels from the handpiece head for at least 2 seconds. (Fig. 12)
CAUTION
When applying spray be sure to hold the handpiece firmly to prevent the handpiece from slipping out
of the hand due to the spray pressure.
Hold the spray can upright.
NSK Care3 Plus
When using NSK Care3 Plus automatic handpiece cleaning and lubrication system, refer to the Care3
System instructions.
4-3 Sterilization
Sterilize the product by autoclave sterilization. Remove the bur after each patient and sterilize as noted
below.
1) Insert into an autoclave pouch. Seal the pouch.
2) Autoclavable under the conditions below.