Allowable environmental conditions
Clean the product after contact with humidity/chemicals/solids, in order to
avoid increased wear and damage (see Page 18).
Prohibited environmental conditions
Solids: highly hygroscopic particles (e.g. talcum), dust in high concentrations
(e.g. construction site), intensive contact with sand
Chemicals/liquids: salt water, perspiration, urine, acids, continuous use in liquid
media

2.4 Service life

This product has been tested by the manufacturer with 2 million load cycles
according to ISO 22675. Depending on the patient's activity level, this cor­
responds to a service life of 2 to 3 years.
3 Safety
3.1 Explanation of warning symbols
Warning regarding possible risks of accident or injury.
CAUTION
Warning regarding possible technical damage.
NOTICE
3.2 General safety instructions
CAUTION
Exceeding the service life and reuse on another patient
Risk of injury due to loss of functionality as well as damage to the product
► Ensure that the approved service life is not exceeded.
► Only use the product on a single patient.
CAUTION
Excessive strain on the product
Risk of injury due to breakage of load-bearing components
► Use the product according to the specified area of application (see
Page 12).
CAUTION
Unallowable combination of prosthetic components
Risk of injury due to breakage or deformation of the product
► Only combine the product with prosthetic components that are
approved for that purpose.