Figure 12
The set point output stays at 0 Vdc until the frequency becomes
higher than 505 Hz (that is 500 Hz + 1 % of 500 Hz), then the output
goes at 24 Vdc and stays at 24 Vdc until the frequency becomes lower
than 495 Hz (that is 500 Hz – 1 % of 500 Hz).
It is possible to delay the set point checking for a programmable
delay time.
The PROGRAMMABLE SET POINT signal has the following default
settings:
The Navigator Software (optional) allows the operator to set all the
NOTE
programmable feature.
When no external input-output device is available this connector
must be closed with the supplied mating connector that short-circuits
the START and INTERLOCK inputs with the GROUND input (see the
following figure).
Special Turbo-V 1K-G User Manual / 87-901-006-01 (A)
reference quantity: frequency
thereshold: 867 Hz
hysteresis: 2 %
activation type: high level
delay time: 0 second
Technical Information
J1 - In-Out
221/258