Window-Meanings

Tab. 7
N.
000
008
100
101
102
103
104
105
106
107
Special Turbo-V 1K-G User Manual / 87-901-006-01 (A)
Read/
Data
Description
Write
Type
R/W
L
Start/Stop
(in remote mode the
window is a read only)
R/W
L
Remote
(default) or Serial
configuration
R/W
L
Soft Start
(write only in Stop
condition)
R/W
N
Set Point type
R/W
N
Set Point threshold
(expressed in Hz, mA or
s)
R/W
N
Set Point delay: time
between the pump
start and the set point
check (seconds)
R/W
L
Set Point signal
activation type: the
signal can be "high
level active" or "low
level active"
R/W
N
Set point hysteresis
(in % of threshold)
R/W
L
Intercooling
R/W
L
Active Stop
(write only in stop)
Technical Information

Window-Meanings

Admitted Values
Start = 1
Stop = 0
Remote = 1
Serial = 0
(default = 1)
YES = 1
NO = 0
0 = Frequency
1 = Current
2 = Time
(default = 0)
(default = 582)
0 to 99999
(default = 0)
0 = high level active
1 = low level active
(default = 0)
0 to 100
(default = 2)
0 = NO
1 = YES
0 = NO
1 = YES