Herunterladen Inhalt Inhalt Diese Seite drucken

Remote Control Mode Operation - Agilent Technologies Turbo-V 300 ICE serie Bedienungshandbuch

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 25
17
Technical Information
Use
Then the Turbo-V ICE pump is slowed down in pro- portion to the gas
load at least until it reaches about 4 or 3 KRPM. As soon as the gas
load decreases, the pump will automatically accelerate to reach
normal operation. The pump can be stopped at any rotational speed
and can be restarted at any rotational speed from either the front
panel buttons or the remote connections. The controller
automatically synchronizes the output to the rotational speed of the
pump and then accelerates linearly up to the nominal speed or
within steps if the Soft Start has been selected.
The display shows:
P
1
until normal operation achieved.

Remote Control Mode Operation

If remote signals are used to operate the controller, it must be
programmed for remote operation (see paragraph "Operating
parameter selections") and when ready to start, the display shows:
R
S
If the Soft Start has been deselected the display shows:
P
With or without Soft Start mode selected the START/STOP and LOW
SPEED front panel push- buttons are inoperative, while the CYCLE
NUMBER and PUMP CURRENT pushbuttons are always active.
250/260
U
M
P
I
2
X
X
E
A
D
Y
O
F
T
S
U
M
P
R
R
E
M
O
T
Turbo-V 300 ICE Controller User Manual / 87-900-902-01 (M)
S
S
T
A
K
R
P
M
F
O
R
R
T
A
R
T
E
A
D
Y
:
E
S
T
A
R
T
I
N
G
H
J
H
S
E
M
O
T
E
H
S
U
S
E
R
T
H
S

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis