Connection J7 – Accessories and Options
Interconnections
Figure 13
The output voltage of the connector J7 is 120 Vac and the maximum current is 1A.
CAUTION!
Do not connect this plug to a device with a leakage current greater than 1.75 mA.
216/260
START S2 signal 24 V, 60 mA, optically isolated output (pin 12 positive, pin 4
negative). The output Voltage will be present when the START push-button on
front panel is pressed or the remote start is present. or the function has been
requested by RS 232/422/485, until NORMAL operation is reached.
OVERTEMP. signal 24 V, 60 mA, optically isolated output (pin 13 positive pin 5
negative). The output Voltage will be present when a fault condition is
displayed on the front panel display, with only in an overtemp condition
FAULT signal 24 V, 60 mA, optically isolated output (pin 13 positive pin 6
negative). The output Voltage will be present when a fault condition is
displayed on the front panel display, with exception of an overtemp condition
Analog output voltage (0 – 10 V) of pump speed (pin 7 positive, pin 14
negative).
Analog output Voltage of DC current drawn by the turbopump (pin 8 positive,
pin 15 negative). 2 Vdc proportional to 1A.
Flow meter alarm, the flow value remained below the threshold for 10 seconds
or more.
Open collector type contacts
J7 connectors
Turbo-V 300 ICE Controller User Manual / 87-900-902-01 (M)