The 120 Vac, 1 A maximum (independent of line Voltage) output
Voltage is present when the main power switch is set to 1 position
and after START push-button is pressed and will remain present
until a fault condition is displayed on the front panel display or the
turbopump is stopped.
The connector J7 is for vent device, forepump connector coils, pump
cooling fan etc. To make connections, remove the plug and wire the
pins (maximum wire size 18 AWG, 1 mm²) as indicated in the figure
to obtain the desired capability.
Forepump relay coil is an independent user supplied item.
NOTE
Connection J10 - Controller to pump Connection
A five-meter long cable is provided to connect the controller to the
pump. The following figures show the controller output connector
configuration where pins:
Figure 14
Turbo-V 300 ICE Controller User Manual / 87-900-902-01 (M)
A-F = upper bearing sensor
B-C-D = 80 Vac 3-phase output to pump motor stator
E = ground
Controller-to-pump connector
(applicable to model 969-9433 and 969-9533)
Technical Information
Interconnections
217/260
17