All the logic output from the controller must be connected at P2
mating connector. With the optional P2 mating connector make the
connection with AWG 24 (0.25 mm2 ) or smaller wire to the pins
indicated in the figure to obtain the desired capability. The following
table describes the signals available on the connector.
The following figure shows a typical logic output connection (relay
coil) but any other device may be connected e.g. a LED, a computer,
etc., and the related simplified circuit of the controller.
Figure 12
Tab. 4
PIN
1-10*
2-10*
3-11*
Turbo-V 300 ICE Controller User Manual / 87-900-902-01 (M)
15
9
Typical output connection open collector
DESCRIPTION
Normal S1 signal 24 V 60 mA, optically isolated output (pin 10 positive, pin 1
negative). The output Voltage will be present when the rotational speed of the
pump is higher than the selected high speed threshold.
Normal S2 signal 24 V 60 mA, optically isolated output (pin 10 positive, pin 2
negative). The output Voltage will be present when the rotational speed of the
pump is higher than the selected low speed threshold.
START S1 signal 24 V, 60 mA, optically isolated output (pin 11 positive, pin 3
negative). The output Voltage will be present when the START push-button on
front panel is pressed or the remote start is present. or the function has been
requested by RS 232/422/485, until NORMAL operation is reached.
Technical Information
8
8
15
9
1
1
17
Interconnections
+ 24