Herunterladen Inhalt Inhalt Diese Seite drucken

Agilent Technologies Turbo-V 300 ICE serie Bedienungshandbuch Seite 213

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 25
Tab. 3
PIN
1-6
2-7
3-8
4-9
In the presence of both S1 and S2 signals, the one that reaches the processor
NOTE
first is executed.
Pin 3-8 must be shorted to allow the Turbo-V 300 ICE pump to start if no
NOTE
interlock contact is connected. If after starting the pump, the interlock contact
opens, it has no effect on the operation and the pump continues to turn.
Turbo-V 300 ICE Controller User Manual / 87-900-902-01 (M)
DESCRIPTION
Remote START/STOP S1 optically isolated from the internal circuit, requires a
permanently closed contact (relay contact. transistor, etc.). When the contact
closes the turbopump starts at high speed and when the contact opens, the
turbopump is stopped. With the remote mode operation selected, the front
panel push-button is inoperative.
Remote START/STOP S2 optically isolated from the internal circuit, requires a
permanently closed contact (relay contact. transistor, etc.). When the contact
closes the turbopump starts at low speed and when the contact opens, the
turbopump is stopped. With the remote mode operation selected, the front
panel push-button is inoperative.
INTERLOCK optically isolated from the internal circuit, this signal can be used
to delay the starting of the turbopump. It requires a permanent closed contact
before starting the turbopump.
SPARE.
Technical Information
Interconnections
213/260
17

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis