4. Push downwards on top shaft to allow top shaft to swing open (Note: top shaft is designed to pivot and not
separate from handle)
REPROCESSING INSTRUCTIONS – Reusable Tips for the Symmetry Sharp Kerrison
reusable tips) and Symmetry Sharp Kerrison
WARNINGS:
 Do not allow the reusable disassembled tips or handle to soak in a saline bath as this may promote
premature corrosion.
The reusable tips must be disassembled from the handle prior to reprocessing and sterilization. The handle
and tips must be reprocessed and sterilized separately once they have been disassembled.
Limitations of Reprocessing
End of life for stainless steel or other metal surgical instruments is normally determined by wear and damage due to
the intended use or misuse, and not reprocessing.
Point of Use Pre-cleaning
Remove excess biologic soil and tissue from the handle and reusable tips using disposable wipes.
As soon as possible after use, set reusable tips and handle in a basin of distilled water or in a tray covered with
damp towels.
LCN 206835-001 Rev I
®
Handle (also known as handle) Only.
®
Handle (also known as
Page 7 of 15