Setting up the Agilent 2100 Bioanalyzer
To set- up the Agilent 2100 Bioanalyzer instrument:
1 Locate bench space that meets the requirements described in
2 Place the instrument on the bench in an upright position.
3 Open the lid of the bioanalyzer and slide the electrode or pressure
Damage by electrostatic discharge
C A U T I O N
The heater plate and electrodes are sensitive to electrostatic discharge.
➔
space"
on page 77.
cartridge into the lid and move the lever into a flat (closed) position.
Adjust the chip selector accordingly (electrode cartridge: position1;
pressure cartridge: position2).
Make sure that you are grounded before inserting a chip to prevent the Agilent 2100
Bioanalyzer from being damaged by electrostatic discharge. The lid is grounded to
leak off electrostatic energy of the operator.
Installing the Agilent 2100 Bioanalyzer System
Setting up the Agilent 2100 Bioanalyzer
4
"Bench
87