Safety Information

Laser Safety

Laser Safety
The Agilent 2100 Bioanalyzer is classified as a "Laser Class 1" product
(IEC825- 1, CFR1040.10). During normal operation of the Agilent 2100
Bioanalyzer no laser light is accessible to the user.
When the laser is in use, the laser light source emits light at a power that
may be harmful to the eyes. To prevent operator exposure to potentially
harmful laser light, an interlock mechanism turns off the laser whenever
the lid of the Agilent 2100 Bioanalyzer is not closed. Furthermore, the lid
can't be completely opened and has a black light absorbing surface in
order to prevent any potential laser light reflection.
Harmful laser light
WA R N I N G
The laser light source emits light at a power that may be harmful to the eyes.
10
Never look into the beam or direct it towards someone else.
System interlocks should never be disabled.