When starting the Agilent 2100 Bioanalyzer software, wait a few seconds
until the self- test procedures for the Agilent 2100 Bioanalyzer are
finished. Otherwise the software may not recognize the Agilent 2100
Bioanalyzer.
Condensation within the instrument
C A U T I O N
Temperature fluctuations could cause condensation within the instrument.
Do not use the Agilent 2100 Analyzer under conditions with temperature
fluctuations.
Let the instrument equilibrate to room temperature for one day to avoid damage of
the electronics.