5.4 EN/IEC 60601 - 1-2 :2007 Sub-clause 5.2.2.2 Table 6
Recommended separation distances between portable and mobile RF communications equipment and the Biosonic
The Biosonic
US100R unit is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer
®
or the user of the Biosonic
US100R unit can help prevent electromagnetic interference by maintaining a minimum distance between the portable
®
and mobile RF communications equipment (transmitters) and the Biosonic
put power of the communications equipment.
Rated maximum output power of
transmitter in watts (W)
0.01
0.1
1.0
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
®
Separation distance according to frequency of transmitter in meters (m)
150kHz to 80MHz
d = 1.2√P
.12
.38
1.2
3.8
12
US100R unit as recommended below, according to the maximum out-
80MHz to 800MHz
d = 1.2√P
.12
.38
1.2
3.8
12
BioSonic
Ultrasonic Scaler
®
US100R unit
®
800MHz to 2.5GHz
d = 2.3√P
.23
.73
2.3
7.3
2.3