Owner's Guide
5.3 EN/IEC 60601 - 1 -2:2007 Sub-clause 5.2.2.2 Table 4
The Biosonic
US100R unit is intended for use in the electromagnetic environment specified below. The customer or the end user of the
®
Biosonic
US100R unit should assure it is used in such an environment.
®
Immunity Test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from objects,
structures and people.
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio,
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which
the Biosonic
US100R unit is used exceeds the applicable RF compliance level above, the Biosonic
®
verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the
Biosonic
US100R unit.
®
b Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
IEC60601 test level
3Vrms
150kHz to 80MHz
3V/m
80MHz to 2.5GHz
Compliance Level
3Vrms
150kHz to 80MHz
3V/m
80MHz to 2.5GHz
Intended Electromagnetic Environment
Portable and mobile RF communications equip-
ment should be used no closer to any part of the
Biosonic
US100R unit, including cables, than the
®
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.2√P
d = 1.2√P
80MHz to 800 MHz
d = 2.3√P
800MHz to 2.5GHz
where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended minimum
separation distance in meters (m).
Field strengths from fixed RF transmitters, as deter-
mined by an electromagnetic site survey a, should
be less than the compliance level in each frequency
range b.
Interference may occur in the vicinity of equipment
marked with the following symbol:
US100R unit should be observed to
®
11