Cleaning and Disinfecting Instructions

Prior to cleaning and surface disinfection, the unit should be turned OFF and disconnected from its
power source.
Cleaning:
Thoroughly wipe all exterior surfaces with a lint-free cloth that has been dampened with an acceptable
cleaning agent. Acceptable cleaning materials are listed below. Remove residual detergent by wiping all
exterior surfaces with a lint-free cloth dampened with distilled water.
Disinfecting:
Disinfect the unit by wiping all exterior surfaces with a lint-free cloth dampened with 80% Ethyl Alcohol.
Allow the unit to air dry.
Cautions:
Do not allow liquids to enter the interior of the unit, and do not permit exterior surfaces to come into
contact with unacceptable solvents such as those listed below, as severe damage to the unit may result.
Acceptable Cleaning Materials:
Vinegar (distilled white vinegar, 5% acidity)
Ammonia-based glass cleaner
Acceptable Disinfecting Material:
Ethanol 80 % by volume
Unacceptable solvents:
MEK (Methyl Ethyl Ketone)
Toluene
Acetone
*Note:
The acceptable cleaning and disinfecting materials listed above have been tested on NDS products and,