3. Remove dummy 2 and insert dummy 3 (Fig. 17).
4. Grind lamination to create hole (Fig. 18).
5. Remove dummy 3 (Fig. 19).
6. Insert magnetic charge port securing in position with silicone adhesive.
Quick Wrist Disconnect Assembly
Insert the coaxial plug into the coupling piece and secure in position using the small "O" ring (Fig. 20).
Insert the coupling piece into the lamination ring and turn until seated. Insert the coupling piece "O" ring and use
the quick wrist disconnect (QWD) release tool to secure in position.
Electrode and battery cables should be inserted into the coaxial plug and the plastic screw used to secure.
To connect the i-Limb hand to the prosthesis:
1. Ensure the i-Limb hand is switched off.
2. Align the hand with the wrist.
3. Give a slight push together.
4. Rotate the hand slightly to ensure engagement.
To remove the i-Limb hand from the prosthesis:
1. Rotate the device through 360° in either direction until it disengages.
2. Remove disengaged hand from the prosthesis.
Wrist Disarticulation Assembly
During fabrication, consideration should be given to create sufficient space for the wiring and switch block and
battery location.
The lamination plate is fabricated directly into the outer socket. No rotation is possible at the wrist and therefore
the fixed position should be set to optimal for the user.
To connect the i-Limb hand to the lamination plate:
1. Remove the palmar fairing of the hand using a T10 screwdriver
2. Feed the power cable through the hole in the plate
3. Align the slots and slide the hand onto the plate
4. Screw the hand and plate together using a T10 screw
5. Replace the palmar fairing, ensuring not to pinch the wiring
6. To remove the i-Limb hand from the lamination plate, complete the steps in reverse.
7. To complete assembly of the prosthesis the battery and electrode cables and hand power cable are connected
to the black WD switch block connector (Fig. 21).
POWER
A fully charged device can be used for up to 16 hours, depending on use.
To charge the i-Limb hand, remove it first from the residual limb. Depending on model, turn either off or to
standby by moving power switch to the left position (Figs. 22, 23).
Magnetic Charge Port
The magnetic charging port allows the battery to be charged, the power switched on/off and the remaining battery
level to be monitored in one place (Fig. 24).
If your i-Limb has an internal battery and magnetic charger port, please follow these steps:
Turning the Device On and Off
– Press switch on charge port for 1 second and release to turn on/off.
– When turned on: the display will fully light up and then dim to a low level.
– When turned off: the display will fully light up and then switch off.