Quick Setup Guide TDL GPro™ 500
Action
– This is the initial message after Power on.
– Wait for the GPro™ 500 to fully boot.
– Check if the GPro™ 500 is powered and
wait until the system is fully started.
– Check the RS485 wiring of the
GPro™ 500 to the M400
– Check with the MT-TDL software and the
Ethernet port if the system is running
correctly.
– If timeout still occurs after 60 s, send unit
back to METTLER TOLEDO.
Send unit back to METTLER TOLEDO
Send unit back to METTLER TOLEDO
Clean corner cube and process window.
Check the gasket between TDL and probe.
Rotate TDL on the probe to maximize Trans-
mission.
Reduce the dustload in the process.
Perform a calibration with the calibration
tube.
If still not successful, send unit back to
METTLER TOLEDO for Flashcard exchange.
Check external pressure sensor and mapping TDL
Check external temperature sensor and
mapping