Standard counting scale
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

STANDARD COUNTING SCALE
Press the key
, the scale will start capturing span and counting down from
5 to 0. Then scale will automatically step forward to the next block.
If "yes" in S2.5 is chosen, the scale needs two points for calibrating the scale.
The procedure to add weight will be as follows (take 5lb scale as example):
After capturing the zero, it will show:
Weight
Data
S 2 6
A D D
>0<
lb
kg
g
Net
ID
APW
B/G
T
Apply test weight (35%FS~65%FS) on the platter and then press the key
The display reads as follows:
Weight
Data
S 2 6
>0<
lb
kg
g
Net
ID
APW
B/G
T
Input a weight data.
Weight
Data
S 2 6
>0<
lb
kg
g
Net
ID
APW
B/G
T
Press the key
, the scale will start capturing span and counting down from
5 to 0. Then scale will automatically step forward to the next step and show:
Weight
Data
S 2 6
A D D
>0<
lb
kg
g
Net
ID
APW
B/G
T
Apply test weight (>85%FS) on the platter and then press the key
display reads as follows:
Weight
Data
S 2 6
>0<
lb
kg
g
Net
ID
APW
B/G
T
Input a weight data.
Weight
Data
S 2 6
>0<
lb
kg
g
Net
ID
APW
B/G
T
Press the key
, the scale will start capturing span and counting down from
5 to 0. Then scale will automatically step forward to the next block
Note: If no calibration is done, the changes on the calibration weight and
capacity will be ignored.
Count
H A L F
M
<
>
PCS
PT
%Accy
Count
0
H A L F
PT
%Accy
M
<
>
PCS
Count
2
h a l f
PT
%Accy
M
<
>
PCS
Count
F U L L
M
<
>
PCS
PT
%Accy
Count
0
F U L L
M
<
>
PCS
PT
%Accy
Count
5
f u l l
PT
%Accy
M
<
>
PCS
27
.