Tab. 5.1:
Coupling plates of the WaveInjector
coupling plate
WI-400
transducer type
M, P, Q
dimensions
After the parameter input, the main menu is displayed.
5.2

Output of the installation parameters

Start a measurement to display the installation parameters of the WaveInjector.
• Select the program branch Measuring.
• Press ENTER.
Measuring\Sound Path
• Enter the number of sound paths, see operating instruction FLUXUS.
• Press ENTER.
The installation parameters of the WaveInjector are displayed, see Fig. 5.1.
• Scroll through the scroll list and write down the following values:
display
Transd. dist.
WI Clamp dist.
WI Plate dist.
WI Transd.pos.
The installation parameters WI Clamp dist. and WI Plate dist. are auxiliary quantities. The transducer distance
Transd. dist. is calculated with these quantities.
For the positioning of the transducers in the transducer box, see section 4.4.
2018-01-19, SUFLUXUS_WIV4-3EN
242.5 mm
description
calculated transducer distance
calculated distance between:
• chain support and jaw (WI-400x-xx-C in diagonal arrangement)
• jaw and jaw (WI-400x-xx-C in reflection arrangement)
• pipe mounting fixture and pipe mounting fixture (WI-400x-xx-T)
calculated distance between the coupling plates
calculated transducer position on the coupling plate (only applicable to the WaveInjector without
the transducer box)
WI-400KG
K, G
279 mm
WaveInjector WI-400
93