Positioning of the transducer in the covers
For the positioning of the transducers in the cover, it is necessary to determine the scale value k, see Fig. 4.60. The scale
value is determined using the following equation:
a l –
k
=
--------- -
2
where
a
transducer distance Transd. dist.
l
distance between the transducer boxes, see Fig. 4.61
k
scale value, see Fig. 4.60
(corresponds to the distance between the edge of the transducer box and the transducer, see Fig. 4.61)
Fig. 4.60:
Positioning of a transducer in the cover
1
0
1
2 Inch
3
mm
10
20 30 40 50 60
scale value k = 23 mm (example)
1 – positioning screw
The scale value is determined as follows:
• For the transducer distance Transd. dist., see section 5.2.
• Measure the distance l, see Fig. 4.61.
• Determine the scale value k using the equation given above.
The transducers are positioned in the covers:
• Loosen the positioning screw of one cover but do not unscrew it completely, see Fig. 4.60.
• Shift the transducer in the cover until the middle of the positioning screw aligns with the scale value k.
• Tighten the positioning screw to fix the transducer in the cover.
• Repeat the steps for the second transducer.