Flexim WaveInjector WI-400 Serie Ergänzung Zur Betriebsanleitung Seite 88

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 4
WaveInjector WI-400
Fig. 4.61:
Positioning of the transducers in the covers (top view)
k
0
1 – positioning screw
Installation of the transducers on the pipe
• Put some coupling foil to the contact surface of the transducer, see Fig. 4.62. The coupling foil can be fixed to the contact
surface with a small amount of coupling compound.
Important!
The transducer cable must not make contact with hot pipes.
• Place the cover with the transducer on the rail, see Fig. 4.62.
Notice!
Make sure that the coupling foil remains on the contact surface of the transducers.
• Tighten the screws of the cover.
Important!
Tighten the screws alternately. Excessive pressure exerted on the transducer can lead to damages.
• Repeat the steps for the second transducer.
88
a
k
l
Supplement to operating instruction - ENGLISH
0
1
2
Inch
3
10
20 30 40 50 60
mm
k
k
2018-01-19, SUFLUXUS_WIV4-3EN