WaveInjector WI-400
4.4

Installation of the transducers

Installation of the transducers in the covers
• Press the transducers firmly into the transducer mounting fixture in the covers until the transducers are tightly fixed, see
Fig. 4.56.
• Only for transducer types K, G and Q:
When mounting, observe the alignment of the transducers in the transducer clamping fixtures, see Fig. 4.57 and Fig.
4.58.
• When mounting, observe the alignment of the transducers in the covers, see Fig. 4.59:
– The arrows on the transducers have to be identical with the arrows on the covers and have to point in the same direction.
– The engravings on the covers form an arrow. The transducer cables show in opposite directions.
Fig. 4.56
Installation of a transducer
1
2
1 – cover
2 – transducer clamping fixture
Fig. 4.57:
Transducer type K, G
hole points away
from transducer
cable
transducer
cable
Fig. 4.59:
Alignment of the transducers in the cover
2
1
1 – transducer cable
2 – arrow end
3 – arrow head
86
Fig. 4.58:
Transducer type Q
hole points toward
transducer cable
transducer
cable
3