Herunterladen Diese Seite drucken

Flexim WaveInjector WI-400 Serie Ergänzung Zur Betriebsanleitung Seite 83

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 4
Supplement to operating instruction - ENGLISH
Installation of the coupling plates on the pipe
• Slide the coupling plate through the coupling plate aperture of the pipe mounting fixture until the middle of the clamping
surface is below the tensioning screw of the pipe mounting fixture, see Fig. 4.50 and Fig. 4.52.
• Fix the coupling plate:
– Tighten the tensioning screw until the disk spring package is almost completely compressed. The metallic coupling foil
adapts to the pipe wall and the coupling plate.
– Loosen the tensioning screw slightly until there is a small gap between the disk springs.
• Push the second coupling plate into the second pipe mounting fixture.
Fig. 4.50: Installed coupling plate (installation with threaded rods, top view)
1 – coupling plate
2 – tensioning screw
Fig. 4.51:
Installed coupling plate (installation with threaded rods, side view)
2018-01-19, SUFLUXUS_WIV4-3EN
2
1
WaveInjector WI-400
83

Quicklinks ausblenden:

Werbung

loading