Herunterladen Diese Seite drucken

Flexim WaveInjector WI-400 Serie Ergänzung Zur Betriebsanleitung Seite 90

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 4
WaveInjector WI-400
Transducer distance check
• Check the transducer distance in the transmitter after approx. 30 minutes, see section 5.2, when the metallic coupling foil
has reached its operating temperature.
• Measure the transducer distance a, see Fig. 4.61:
a = 2 · k + l
where
k
– scale value
l
– distance between the transducer boxes
Notice!
The transducer distance has to be adjusted correctly.
• Correct the transducer distance if the measured transducer distance does not correspond to the calculated transducer
distance Transd. dist., see section 5.2. For changes concerning the transducer distance, see section 4.5.
• Tighten the tensioning screws again, if necessary.
Fig. 4.64:
Installed WaveInjector WI-400x-xx-C
Fig. 4.65:
Installed WaveInjector WI-400x-xx-T